| Chapter 13: Spectroscopy |
Nuclear Magnetic Resonance (NMR) Spectroscopy
Since a nucleus is a charged particle in motion, it will develop a magnetic
field. 1H and 13C have nuclear spins
of 1/2 and so they behave in a similar fashion to a simple, tiny bar magnet.
In the absence of a magnetic field, these are randomly oriented but when
a field is applied they line up parallel to the applied field, either spin
aligned or spin opposed. The more highly populated state is
the lower energy spin state spin aligned situation. Two schematic
representations of these arrangements are shown below:


![]() |
In NMR, EM radiation is used to "flip" the alignment of nuclear spins from the low energy spin aligned state to the higher energy spin opposed state. The energy required for this transition depends on the strength of the applied magnetic field (see below) but in is small and corresponds to the radio frequency range of the EM spectrum. |
| As this diagram shows, the energy required for the spin-flip depends on the magnetic field strength at the nucleus. With no applied field, there is no energy difference between the spin states, but as the field increases so does the separation of energies of the spin states and therefore so does the frequency required to cause the spin-flip, referred to as resonance. | ![]() |
![]() |
The basic arrangement of an NMR spectrometer is shown to the left. The sample is positioned in the magnetic field and excited via pulsations in the radio frequency input circuit. The realigned magnetic fields induce a radio signal in the output circuit which is used to generate the output signal. Fourier analysis of the complex output produces the actual spectrum.The pulse is repeated as many times as necessary to allow the signals to be identified from the background noise. |